Deklene® II Monofilament, Non-absorbable Suture

Choose strong and smooth Deklene® suture for the most delicate of specialty procedures.

Deklene Image

Material: Polypropylene (PP)


Coating: None


Characteristics: Deklene® II suture provides outstanding handling, accurate knot placement, and knot security. The soft, supple suture material reduces memory and improves handling.


Indications: Deklene® II suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures, but not for use in ophthalmic procedures.


Sizes: USP 2 through 9-0


Construction: Monofilament


Color:

  • Blue (colorant: [phthalocyaninato (-2)] copper)
  • Clear

Sterilization: Ethylene oxide (EtO)


Configurations:

  • Non-sterile spools and cut-lengths
  • Sterile single peel and straight packs

Custom Configurations are Available

Deklene® II can be customized to meet the requirements of your unique application, or incorporated into an anchor or driver of your design. Teleflex Medical OEM is a single-source, product development partner that can assemble, package, sterilize, and private label the final product.


Deklene® II Suture and Needle Combinations

Teleflex Medical OEM can help you determine a configuration suited for a particular surgical application. Then we will manufacture, sterilize, package, and private label your product with attention to quality every step of the way. More »


USP Requirements for Non-absorbable Surgical Suture


Suture Size USP Diameter
(mils.)
USP Knot Pull
(lbs.)
USP METRIC MIN MAX MIN
9-0 0.3 1.2 1.5 0.095
8-0 0.4 1.6 2.3 0.13
7-0 0.5 2.0 3.3 0.24
6-0 0.7 2.8 3.9 0.44
5-0 1 3.9 5.9 0.88
4-0 1.5 5.9 7.8 1.32
3-0 2 7.9 9.8 2.12
2-0 3 11.8 13.3 3.18
0 3.5 13.8 15.7 4.76
1 4.0 15.7 19.6 6.0
2 5 19.7 23.6 7.76

Please note: Specifications in the chart are the United States Pharmacopeia (USP) requirements. Deklene® II suture meets all the requirements established by the USP for non-absorbable surgical suture, except for suture sizes 7-0 and 8-0 which differ from the USP in diameter only. Knot pull values are for sterile suture.